Abstract

Digital radiography overcomes many of the limitations of the traditional screen/film system. Further enhancements in the digital radiography image are possible if the X-ray image receptor could measure the energy of individual photons instead of simply integrating their energy, as is the case at present. A prototype photon counting scanned projection radiography system has been constructed, which combines a Gas Electron Multiplier (GEM) and a Gas Microstrip Detector (GMD) using Xe : CH 4 (90 : 10) at high pressure. With the gain contribution from the GEM, the GMD can be operated at lower and safer voltages making the imaging system more reliable. Good energy resolution, and spatial resolution comparable to that of screen/film, have been demonstrated for the GEM/GMD hybrid imaging system in photon counting mode for X-ray spectra up to 50 kV.