Abstract

As a reference photon field, several radionuclides have been used frequently, such as 241Am,137Cs and60 Co for calibration. These nuclides provide mono-energy photons for dosemeters covering few tens of keV-MeV. The main energy around 200keV is important for both environmental and medical fields since the former should consider scattering photons and the later should measure photons from X-ray generator. In our previous work, a backscattered layout can provide a uniform photon field spectra and dose rate with an energy of 190keV by using an affordable intensity 137 Cs gamma source. Several other quasi-monoenergetic photon fields in the range of 100-200keV could be obtained by using several available gamma sources. Two calibrated environmental CsI(Tl) survey meters, Horiba PA-1000 and Mr. Gamma A2700, had been measured with the developed backscattered photon field to understand energy-dependent features in order to confirm dosemeter readings. Consequently, both scintillator instruments are sensitive for measurements of the relatively low dose rates at 190keV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.