Abstract
As a reference photon field, several radionuclides have been used frequently, such as 241Am,137Cs and60 Co for calibration. These nuclides provide mono-energy photons for dosemeters covering few tens of keV-MeV. The main energy around 200keV is important for both environmental and medical fields since the former should consider scattering photons and the later should measure photons from X-ray generator. In our previous work, a backscattered layout can provide a uniform photon field spectra and dose rate with an energy of 190keV by using an affordable intensity 137 Cs gamma source. Several other quasi-monoenergetic photon fields in the range of 100-200keV could be obtained by using several available gamma sources. Two calibrated environmental CsI(Tl) survey meters, Horiba PA-1000 and Mr. Gamma A2700, had been measured with the developed backscattered photon field to understand energy-dependent features in order to confirm dosemeter readings. Consequently, both scintillator instruments are sensitive for measurements of the relatively low dose rates at 190keV.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.