Abstract

Abstract We measure the ability of the Leonardo Selex Advanced Photodiode HgCdTe Infrared Array (SAPHIRA) avalanche photodiode array to perform photon counting. The current SAPHIRA arrays achieve >90% single-photon efficiency (independent of quantum efficiency expected to be also >90%) and a time resolution of 125 μs with a dark current of 21e − s−1 pix−1. Our characterization of several iterations of the SAPHIRA detector over the past 3 years of its development have also revealed a broader pulse height distribution than was originally expected.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.