Abstract
The aim of this study was to assess in-stent lumen visibility and quantitative image characteristics of different coronary stents using a novel photon-counting detector (PCD) computed tomography (CT) system in comparison to a state-of-the-art energy-integrating detector (EID) CT scanner. In this in vitro phantom study, 28 different coronary stents ranging from 2.25 to 4.5 mm lumen diameter were expanded into plastic tubes filled with contrast agent. Stent-containing plastic tubes were positioned in a custom-made emulsion-filled phantom, which was inserted into an anthropomorphic phantom simulating a medium-sized patient. Computed tomography scans were acquired parallel to the scanners' z axis using a novel cadmium telluride-based PCD CT system (SOMATOM CountPlus; Siemens Healthcare GmbH, Forchheim Germany), operating in 2 different modes (standard-resolution mode [SR] and ultra-high-resolution [UHR] mode), and a latest generation dual-source EID CT system (SOMATOM Force; Siemens Healthcare GmbH, Forchheim). CTDIvol-matched images were reconstructed with comparable convolution kernels and using the same reconstruction parameters. In-stent lumen visibility (in %), increase in in-stent attenuation (expressed as Δ in-stent CT attenuation), and image noise (in Hounsfield unit) were manually measured. Parts of the image analysis (in-stent lumen visibility) were additionally performed in an automated way. Differences were tested using Wilcoxon signed rank test. The best in-stent lumen visibility was achieved with the PCD-UHR mode and the lowest noise levels with the PCD-SR mode. The median in-stent lumen visibility was significantly higher (P < 0.001) with PCD (SR, 66.7%; interquartile range [IQR], 63.3-72.3; UHR, 68.9%; IQR, 64.4-74.4) compared with EID (65.4%; IQR, 62.2-70.4). The Δ in-stent CT attenuation was significantly lower for PCD in both SR (78 HU; IQR, 46-108; P = 0.024) and UHR (85 HU; IQR, 59-113; P = 0.006) compared with EID (108 HU; IQR, 85-126). Image noise was significantly lower (P < 0.001) for PCD-SR (21 HU; IQR, 21-21) compared with EID images (25 HU; IQR, 24-25.0). The PCD provides superior in-stent lumen visibility and quantitative image characteristics when compared with conventional EID.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.