Abstract
The beam quality correction factor kQ,Q0 and the perturbation factor pQ for photon beams were calculated for the NE2571A and NE2581A ionization chambers, using the Monte Carlo simulation code PENELOPE. Results are compared to those quoted for the NE2571 and NE2581 chambers in previous works. Both kQ,Q0 and pQ obtained for NE2571A and NE2581A chambers agree with those of their predecessors NE2571 and NE2581 ones.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.