Abstract

Background & purposeTo evaluate treatment planning system (TPS) beam modeling parameters as contributing factors to IMRT audit performance. Materials & methodsWe retrospectively analyzed IROC Houston phantom audit performance and concurrent beam modeling survey responses from 337 irradiations performed between August 2017 and November 2019. Irradiation results were grouped based on the reporting of typical or atypical beam modeling parameter survey responses (<10th or >90th percentile values), and compared for passing versus failing (>7% error) or “poor” (>5% error) irradiation status. Additionally, we assessed the impact on the planned dose distribution from variations in modeling parameter value. Finally, we estimated the overall impact of beam modeling parameter variance on dose calculations, based on reported community variations. ResultsUse of atypical modeling parameters were more frequently seen with failing phantom audit results (p = 0.01). Most pronounced was for Eclipse AAA users, where phantom irradiations with atypical values of dosimetric leaf gap (DLG) showed a greater incidence of both poor-performing (p = 0.048) and failing phantom audits (p = 0.014); and in general, DLG value was correlated with dose calculation accuracy (r = 0.397, p < 0.001). Manipulating TPS parameters induced systematic changes in planned dose distributions which were consistent with prior observations of how failures manifest. Dose change estimations based on these dose calculations agreed well with true dosimetric errors identified. ConclusionAtypical TPS beam modeling parameters are associated with failing phantom audits. This is identified as an important factor contributing to the observed failing phantom results, and highlights the need for accurate beam modeling.

Highlights

  • Background & purposeTo evaluate treatment planning system (TPS) beam modeling parameters as contributing factors to IMRT audit performance

  • We reviewed 337 IMRT phantom audits from IROC Houston between August 2017 to November 2019, along with concurrently reported TPS beam modeling parameters associated with each irradiation

  • We evaluated the impact of atypical beam modeling parameter values on the dose distribution in IROC phantoms, in particular, the dose profiles through the target volumes

Read more

Summary

Introduction

To evaluate treatment planning system (TPS) beam modeling parameters as contributing factors to IMRT audit performance. Irradiation results were grouped based on the reporting of typical or atypical beam modeling parameter survey responses (90th percentile values), and compared for passing versus failing (>7% error) or ‘‘poor” (>5% error) irradiation status. We assessed the impact on the planned dose distribution from variations in modeling parameter value. We estimated the overall impact of beam modeling parameter variance on dose calculations, based on reported community variations. Conclusion: Atypical TPS beam modeling parameters are associated with failing phantom audits. This is identified as an important factor contributing to the observed failing phantom results, and highlights the need for accurate beam modeling.

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.