Abstract

A thermoluminescent dosemeter (TLD) mailed dose audit programme was performed at five radiotherapy clinics in Turkey. The intercomparison was organised by the University of Wisconsin Radiation Calibration Laboratory (UWRCL), which was responsible for the technical aspects of the study including reference irradiations, distribution, collection and evaluation. The purpose of these audits was to perform an independent dosimetry check of the radiation beams using TLDs sent by mail. Acrylic holders, each with five TLD chips inside and instructions for their irradiation to specified absorbed dose to water of 2 Gy, were mailed to all participating clinics. TLD irradiations were performed with a 6 MV linear accelerator and (60)Co photon beams. The deviations from the TL readings of UWRCL were calculated. Discrepancies inside the limits of ±5 % between the participant-stated dose, and the TLD-measured dose were considered acceptable. One out of 10 beams checked was outside this limit, with a difference of 5.8 %.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.