Abstract

The internal standard method coupled with the standard addition method has been applied to photon activation analysis and proton activation analysis of minor elements and trace impurities in various types of iron and steel samples issued by the Iron and Steel Institute of Japan (ISIJ). Samples and standard addition samples were once dissolved to mix homogeneously, an internal standard and elements to be determined and solidified as a silica-gel to make a similar matrix composition and geometry. Cerium and yttrium were used as an internal standard in photon and proton activation, respectively. In photon activation, 20 MeV electron beam was used for bremsstrahlung irradiation to reduce matrix activity and nuclear interference reactions, and the results were compared with those of 30 MeV irradiation. In proton activation, iron was removed by the MIBK extraction method after dissolving samples to reduce the radioactivity of56Co from iron via56Fe(p,n)56Co reaction. The results of proton and photon activation analysis were in good agreement with the standard values of ISIJ.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call