Abstract

A new procedure is developed for the photometric determination of trace selenium in aqueous solutions. The selection of 2-(p-nitrophenyl)-3,5-diphenyltetrazolium chloride as a reagent for selenium is justified. The proposed sample preparation procedure involves gas extraction of selenium as hydrogen selenide followed by its liquid-adsorption extraction from the gas phase to an aqueous reagent solution with the formation of a water-insoluble formazan. Formazan formed upon the absorption of hydrogen selenide is extracted with isoamyl alcohol. The concentration of selenium is determined from the absorbance of the formazan extract in the isoamyl alcohol. The procedure allows the determination of 10–120 μg/L selenium.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.