Abstract
Photoluminescence and optical transmission spectra of several samples of natural and synthetic diamond and its imitators – fianite and corundum – are investigated. The band-A of luminescence at 440 nm, the vibronic N3 system of luminescence and absorption at 415.2 nm, the fundamental absorption edge at 225 nm, and the secondary absorption below 308 nm are the main identifying markers of natural diamonds. For synthetic diamonds, however, such identifying markers are the free-exciton luminescence at 235 nm, the band-A, and the fundamental absorption edge. Fianites can be identified by the structureless wideband at 500 nm and the wide transmission band in the entire visible range. Colored corundum samples with chrome impurities emit the narrow line at 693 nm and show the absorption band in the 500–600 nm spectral range. A new method for diamond express identification is developed on the basis of measurement of photoluminescence and optical transmission spectra of the samples. It is shown that a diamond tester can be designed combining a spectrometer and a KrCl-excilamp radiating at 222 nm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.