Abstract

Abstract A photoionisation detector (PID) was coupled to packed-capillary liquid and supercritical-fluid chromatography to study its performance. Several mobile phases were tested to evaluate the potential of liquid chromatography with photoionisation detection, LC-PID. The behaviour of the PID was not as good as in gas chromatography (GC), due to the absorption of photons by the mobile phase vapour. Therefore, the minimum detection limits (MDLs) were high compared to those in GC-PID, being at the low nanogram level for, e.g., ketones, aldehydes and amides. Coupling of the PID with supercritical-fluid chromatography (SFC) using modified carbon dioxide gave more satisfactory results. For aromatic compounds like phenanthrene and pyrene MDLs were found to be 12 and 20 pg, respectively. These values were almost the same as found in GC-PID. Separation and detection of aliphatic C4-C13 carboxylic acids and some organosulphur/phosphorus pesticides (disulfoton, ethion and sulfotepp) showed the possibility to dete...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.