Abstract

BackgroundReal-time photoacoustic gas monitoring is used for personnel exposure and environmental monitoring, but its accuracy varies when organic solvents such as alcohol contaminate measurements. This is problematic for anesthetic gas measurements in hospitals, because most disinfectants contain alcohol, which could lead to false-high gas concentrations. We investigated the cross-sensitivities of the photoacoustic gas monitor Innova 1412 (AirTech Instruments, LumaSense, Denmark) against alcohols and alcoholic disinfectants while measuring sevoflurane, desflurane and isoflurane in a laboratory and in hospital during surgery.Methods25 mL ethyl alcohol was distributed on a hotplate. An optical filter for isoflurane was used and the gas monitor measured the ‘isoflurane’ concentration for five minutes with the measuring probe fixed 30 cm above the hotplate. Then, 5 mL isoflurane was added vaporized via an Anesthetic Conserving Device (Sedana Medical, Uppsala, Sweden). After one-hour measurement, 25 mL isopropyl alcohol, N-propanol, and two alcoholic disinfectants were subsequently added, each in combination with 5 mL isoflurane. The same experiment was in turn performed for sevoflurane and desflurane. The practical impact of the cross-sensitivity was investigated on abdominal surgeons who were exposed intraoperatively to sevoflurane. A new approach to overcome the gas monitor’s cross-sensitivity is presented.ResultsCross-sensitivity was observed for all alcohols and its strength characteristic for the tested agent. Simultaneous uses of anesthetic gases and alcohols increased the concentrations and the recovery times significantly, especially while sevoflurane was utilized. Intraoperative measurements revealed mean and maximum sevoflurane concentrations of 0.61 ± 0.26 ppm and 15.27 ± 14.62 ppm. We replaced the cross-sensitivity peaks with the 10th percentile baseline of the anesthetic gas concentration. This reduced mean and maximum concentrations significantly by 37% (p < 0.001) and 86% (p < 0.001), respectively.ConclusionPhotoacoustic gas monitoring is useful to detect lowest anesthetic gases concentrations, but cross-sensitivity caused one third falsely high measured mean gas concentration. One possibility to eliminate these peaks is the recovery time-based baseline approach. Caution should be taken while measuring sevoflurane, since marked cross-sensitivity peaks are to be expected.

Highlights

  • Real-time photoacoustic gas monitoring is used for personnel exposure and environmental monitoring, but its accuracy varies when organic solvents such as alcohol contaminate measurements

  • We investigated the cross-sensitivities of a photoacoustic gas monitor against alcohols and alcoholic disinfectants (AD) while measuring isoflurane (ISO), sevoflurane (SEVO), and desflurane (DES) in the air

  • The decay curves corresponded to recovery times of 52 min and exponential functions with R2 = 0.999 for ISO, R2 = 0.995 for SEVO, and R2 = 0.999 for DES) (n = 5 per volatile anesthetics (VA)) (Fig. 2)

Read more

Summary

Introduction

Real-time photoacoustic gas monitoring is used for personnel exposure and environmental monitoring, but its accuracy varies when organic solvents such as alcohol contaminate measurements This is problematic for anesthetic gas measurements in hospitals, because most disinfectants contain alcohol, which could lead to false-high gas concentrations. The Occupational Safety and Health Administration of the United States recommends the control of VA exposure by engineering controls, good work practices, use of personal protective equipment, administrative controls, and personnel exposure and environmental monitoring. The latter is especially significant, because it demonstrates the effectiveness of the gas control program and should be performed every six months [4]. This is problematic for VA measurements in hospitals, because many commonly used substances contain alcohol, which may lead to a misjudgment of the actual pollution (Table 1)

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.