Abstract
Phoswich detectors for simultaneous counting of α, β and γ rays have been developed: ZnS(Ag)/Au Mylar/NE102A, ZnS(Ag)/Au Mylar/BGO and ZnS(Ag)/NaI(Tl) for α and β(γ) rays and ZnS(Ag)/Au Mylar/NE102A/BGO and ZnS(Ag)/NE102A/NaI(Tl) for α, β and γ rays. They were prepared by coupling a ZnS(Ag) film scintillator for α counting with a scintillator(s) for β and γ counting having different rise time. In order to adjust each component of pulse height within a given dynamic range, a sheet of Au-coated Mylar (Au Mylar) was used, if necessary, as an optical ND filter for lowering transmittance of scintillation of the ZnS(Ag). Characteristics of these phoswiches were examined by a technique of pulse-shape discrimination. Excellent discrimination among the radiations was attained and small tailings from each other peak were obtained for the prepared phoswiches.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.