Abstract
Analysis of diastereomers of phosphorothioate oligonucleotides in ion-pairing reversed-phase liquid chromatography is affected not only by the character and concentration of ion-pairing system, but also by the separation temperature. In this work, eight ion-pairing systems at two concentrations buffered with acetic acid were used with octadecyl column to investigate the effects of temperature (in the range from 20 °C to 90 °C) on retention, diastereomeric separation, resolution of mers of different length and resolution of oligonucleotides with different number of phosphorothioate linkages. It was observed that elevated temperature suppresses the diastereomeric separation and oligonucleotide peaks become narrower. This improves the resolution of n and n-1 mers at elevated temperature. Plots of ln k (k = retention factor) versus reciprocal absolute temperature show that for 100 mM ion-pairing systems the increase in temperature does not lead to simple decrease in oligonucleotides retention as generally observed in reversed-phase liquid chromatography. The aim of this work is to improve chromatographic method for analysis of phosphorothioate oligonucleotides.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.