Abstract

This article describes a simple procedure for the detection of phosphorylated peptides by comparable positive and negative ion mode matrix-assisted laser desorption/ionization mass spectrometry measurements. Based on studies with phosphorylated peptides (EAIXAAPFAK, X = pS, pT, pY) and their corresponding non-phosphorylated analogs, it was found that phosphopeptides, which are characterized by a low ionization efficiency in the positive ion mode, exhibit drastically increased signal intensities in the negative ion mode compared to their non-phosphorylated analogs. The effect was successfully used to identify phosphorylated sequences of the commonly used phosphoprotein standards, protein kinase A and beta-casein, by peptide mass fingerprint analyses of the corresponding Lys C and trypsin digests using both (positive and negative) ion modes. The comparison of positive and negative ion spectra of a given protein digest (relative intensity([M - H]-)/relative intensity([M + H]+)) can be used to identify any phosphopeptides present which may then be separated and analyzed further.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.