Abstract
New MR-techniques for quantitative blood flow registration such as phase-mapping (a two-dimensional space-resolved technique with a time-averaged measurement of blood flow) or RACE (real-time acquisition and evaluation of blood flow in one-dimensional space projection) are available for the diagnosis of valvular heart disease. Initial results of grading aortic valve stenosis by these methods are shown in comparison to continuous wave Ultrasound-Doppler. Two groups of 15 patients were examined by RACE or phase-mapping, 12 respectively 8 of whom suffered from an aortic valve stenosis. The shape of blood flow profiles as well as grading of aortic valve stenosis show high concordance when comparing the results of MR and Doppler technique. Good reliability and practicability of the demonstrated MR-method are shown. With respect to the results of RACE and phase-mapping the development of an alternative and competing MR-method for the evaluation of valvular heart disease and shunt diagnostics seems possible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.