Abstract

We report here on a low-cost, optical oxygen sensor as an attractive alternative to the widely used amperometric Clark-type oxygen electrode for measuring dissolved oxygen tensions in cell cultures and bioreactor. Our sensor is based on the defferential quenching of the fluorescence lifetime of chromophore in response to the partial pressure of oxygen. This is measured as a phase shift in fluorescence emission from the chromophore due to oxygen quenching when excited by an intensity modulated beam of light. In this article we demonstrate the advantages of lifetime-based optical methods over both intensity based optical methods and amperometric electrodes. Our sensor is particularly suitable for measuring dissolved oxygen in bioreactors. It is autoclavable, is free of maintenance requirements, and solvents the problems of long-term stability, calibration drifts, and reliable measurement of low oxygen tensions in dense microbial cultures that limit the utility of Clark-type elcectordes. (c) 1994 John Wiley & Sons, Inc.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.