Abstract

The quadrature sampling technique as a means of detecting the envelope of RF waveform in the baseband is well known. If this technique is applied to a focused ultrasound imaging system using an array transducer, whether it is a synthetic or nonsynthetic focusing system, unwanted phase terms appear in the expressions of the inphase and quadrature components of the baseband signal when an appropriate delay time is introduced to each channel signal for the purpose of focusing. The expressions of the inphase and quadrature components from the point of focusing are derived and analyzed in detail, and a scheme to eliminate the unwanted phase terms is proposed. The resulting phase-error-free quadrature sampling technique is applied to the synthetic focusing system; a system block diagram together with the simulation and experimental results are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.