Abstract

The acoustic holography method was used to characterize a therapeutic focused fully populated 256-element ultrasonic transducer array. Elements of the array with the shape of equal area polygons are densely arranged in an irregular pattern on a spherically concave surface with a radius of curvature of 150 mm and a diameter of 200 mm. The array has a central frequency of 1.2 MHz and is designed to operate in water. The performance of individual array elements was studied based on the holographically reconstructed normal velocity distribution over the array surface. It was shown that with the same electrical signals applied to the elements, their acoustic responses had a phase deviation relative to the nominal values, which can be caused either by the asphericity of the array surface, or by the introduction of additional phase delays by the electrical matching network. To compensate for the detected parasitic phase shifts of the elements and restore the effective sphericity of the radiating surface, the Verasonics V-1 control system was used. The hologram measured after making the correction, as well as the shape of the focal region and acoustic pressure magnitude at the focus, separately measured by a hydrophone, showed that the proposed method reconstructed the nominal operating parameters of the array with high accuracy.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call