Abstract

The quality of focusing in high-frequency ultrasound imaging is significantly affected by the length of the depth of field (DOF). A dual-concentric transducer combined with a phase-apodisation scheme is presented to achieve the extended DOF maintaining signal-to-noise ratio. As a preliminary study, computational simulation by using a Field-II program is conducted to demonstrate the feasibility. A dual-concentric transducer is composed of a disc- and a ring-type element with confocal apertures. When two input signals with 0° and 180° phases are simultaneously applied to the inner and outer elements, a bifocal-zone is generated in the axial direction. The overall − 6 dB DOF is 40% longer compared with a single element transducer. Thus, the proposed scheme can be a potential approach to increase the DOF for high-frequency ultrasound imaging.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.