Abstract
Purpose Hypoxemia during a failed airway scenario is life threatening. A dual-lumen pharyngeal oxygen delivery device (PODD) was developed to fit inside a traditional oropharyngeal airway for undisrupted supraglottic oxygenation and gas analysis during laryngoscopy and intubation. We hypothesized that the PODD would provide oxygen as effectively as high-flow nasal cannula (HFNC) while using lower oxygen flow rates. Methods We compared oxygen delivery of the PODD to HFNC in a preoxygenated, apneic manikin lung that approximated an adult functional residual capacity. Four arms were studied: HFNC at 20 and 60 liters per minute (LPM) oxygen, PODD at 10 LPM oxygen, and a control arm with no oxygen flow after initial preoxygenation. Five randomized 20-minute trials were performed for each arm (20 trials total). Descriptive statistics and analysis of variance were used with statistical significance of P < 0.05. Results Mean oxygen concentrations were statistically different and decreased from 97% as follows: 41 ± 0% for the control, 90 ± 1% for HFNC at 20 LPM, 88 ± 2% for HFNC at 60 LPM, and 97 ± 1% (no change) for the PODD at 10 LPM. Conclusion Oxygen delivery with the PODD maintained oxygen concentration longer than HFNC in this manikin model at lower flow rates than HFNC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.