Abstract

To respond to shortage of pilocarpine discs due to CE-licensing problems a pharmacy compounded pilocarpine HCL solution was developed and validated for use in CF diagnosis.The aim of this study was to compare the results from a pharmacy compounded pilocarpine HCL solution versus Pilogel® discs for the measurements of sweat chloride concentrations.Ten pediatric and adult patients with CF underwent a sweat test using both Pilogel® discs and pilocarpine HCL solution.The average difference between both methods was -3.25 mmol/L (95% Limits of Agreement: -7.19 [95% CI: -9.19;-5.19] and 0.69 [95% CI: -1.31;2.69] mmol/L. Passing-Bablok regression showed that zero was enclosed with the 95% CI of the calculated intercept (0.15 [95% CI: -1.70;1.42] mmol/L).These data show a good agreement in chloride concentrations obtained using the two pilocarpine products. Therefore, the pharmacy compounded pilocarpine HCL solution can be used as an alternative for Pilogel® discs during times of shortage.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.