Abstract

We investigated dose reduction ability of an iterative reconstruction technology for low-dose computed tomography (CT) for lung cancer screening. The Sinogram Affirmed Iterative Reconstruction (SAFIRE) provided in a multi slice CT system, Somatom Definition Flash (Siemens Healthcare) was used. An anthropomorphic chest phantom (N-1, Kyoto Kagaku) was scanned at volume CT dose index (CTDIvol) of 0.50-11.86 mGy with 120 kV. For noise (standard deviation) and contrast-to-noise ratio (CNR) measurements, CTP486 and CTP515 modules in the Catphan (The Phantom Laboratory) were scanned. Radiological technologists were participated in the perceptual comparison. SAFIRE reduced the SD values by approximately 50% compared with filter back projection (FBP). The estimated dose reduction rates by SAFIRE determined from the perceptual comparison was approximately 23%, while 75% dose reduction rate was expected from the SD value reduction of 50%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.