Abstract

A phantom design method suitable for high-field MRI based on the RF field wave characteristics of sample and experimental validations at 7.0 T and 3.0 T are presented. The RF field distribution in a phantom with a given RF coil system is primarily determined by the sample size relative to the wavelength inside the sample, and the ratio between the displacement and conduction currents. Experimental results demonstrate that the MR image intensity patterns associated with wave behavior in human samples at a given field strength can be reproduced with a phantom at the same or different field strengths once the dimension and penetration constant are scaled by the corresponding wavelength in the sample medium.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.