Abstract

In event-related functional MRI, there exist limits on the time length of the experiments on human subjects and the imaging speed. Due to these limitations, data truncation and undersampling have to be used in functional MRI signal acquisition. The effect of these factors on the hemodynamic deconvolution is investigated experimentally and a phantom calibration method to improve the hemodynamic response is developed. It is observed that the high frequency components generated due to data truncation can fold back into low frequencies when the sampling rate is not sufficiently high. This aliasing can introduce significant noise in hemodynamic deconvolution and can reduce the accuracy of the temporal characterization of hemodynamic response. A SMARTPHANTOM™ BOLD simulator is used to calibrate the aliasing effect in an event-related functional MRI experiment. With the calibration, an anti-aliasing method is used to suppress the aliasing and this resulted in an improved temporal characterization of hemodynamic response in event-related fMRI.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.