Abstract

A prompt gamma neutron activation analysis (PGNAA) system has been recently developed at the 30-kW research reactor In-Hospital Neutron Irradiator (IHNI) in Beijing. Neutrons from the specially designed thermal neutron beam were used. The thermal flux of this beam is 3.08×106cm−2s−1 at a full reactor power of 30kW. The PGNAA system consists of an n-type high-purity germanium (HPGe) detector of 40% efficiency, a digital spectrometer, and a shielding part. For both the detector shielding part and the neutron beam shielding part, the inner layer is composed of 6Li2CO3 powder and the outer layer lead. The boron-10 sensitivity of the PGNAA system is approximately 2.5cps/ppm. Two calibration curves were produced for the 1–10ppm and 10–50 ppm samples. The measurement results of the control samples were in accordance with the inductively coupled plasma atomic emission spectroscopy (ICP-AES) results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.