Abstract

ObjectivesTo evaluate prospective single centre experience of mid-term safety and efficacy of perventricular device closure of isolated large muscular ventricular septal defect (mVSD) in high-risk infants. BackgroundSurgical closures of large mVSD in infants represent a challenge with significant morbidity. MethodsBetween August 2008–2010, perventricular closure was attempted in 24 infants of 6.01 ± 2.37 months age and 4.27 ± 0.56 kg weight under TEE guidance. ResultsThe device was successfully deployed in 21/24 infants. Size of mVSD was 8.42 ± 1.46 mm (6.1–12 mm). Mean procedure time was 28.8 ± 11.7 min. The closure rate was 84% immediately and 100% at 6 months. Four patients suffered major complications: 2-died, 1-esophageal perforation, 1-persistent CHB. At 26.23 ± 6.63 months follow-up two patients were symptomatic: 1-required device retrieval, 1-died of severe gastroenteritis. ConclusionPerventricular device closure of isolated mVSD appears feasible option at mid-term follow-up and may either substitute or complement the conventional surgical technique in selected cases depending on institutional paediatric cardiac surgery performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.