Abstract

This paper presents some results of the development of two new scintillation detectors usable for the detection and spectrometry of neutron and photon components of the mixed field. It is a series of liquid organic scintillators containing carbon, hydrogen, nitrogen and oxygen. Elements of individual scintillator designs are arranged in various chemical bonds defining physical and chemical properties of the scintillators. These scintillators have advantageous useful properties. They are non-hygroscopic, not poisonous and are not sensitive to contact with the outside atmosphere and can be stored in conventional laboratory containers.

Highlights

  • The authors list basic properties of two types of new liquid scintillators LSB205 and LSB206 that are compared with the characteristics of the BC 501 A (NE-213) liquid scintillator, which can be considered the standard used for many years for detection and spectrometry in the mixed field of neutron and photon radiation

  • In order to use these scintillators for gamma/neutron mixed field spectrometry, it is necessary to obtain light output function at least for protons

  • The obtained scintillator temperature dependencies described in this article will enable to refine the measurement of the amplitude resolution of the neutron spectrum measured at the collimated reactor output beam moderated by silicon

Read more

Summary

INTRODUCTION

The authors list basic properties of two types of new liquid scintillators LSB205 and LSB206 that are compared with the characteristics of the BC 501 A (NE-213) liquid scintillator, which can be considered the standard used for many years for detection and spectrometry in the mixed field of neutron and photon radiation. This paper mainly contains qualitative and quantitative data on the resolution of neutron and photon particles detected by the scintillator. In order to use these scintillators for gamma/neutron mixed field spectrometry, it is necessary to obtain light output function at least for protons. The obtained scintillator temperature dependencies described in this article will enable to refine the measurement of the amplitude resolution of the neutron spectrum measured at the collimated reactor output beam moderated by silicon. This spectrum contains several neutron lines with well-defined energies which could substitute hard to obtain mono-energetic neutron sources [4], [5].

METHODOLOGY
EXPERIMENTAL SETUP
TEMPERATURE TEST RESULTS
MeV 4 MeV 5 Mev
CONCLUSION
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.