Abstract

The recommendations recently proposed by the International Standards Organization (ISO) and the American Conference of Governmental Industrial Hygienists (ACGIH) on particle-size definitions have stressed the importance of gathering information on a larger size interval than previously used and, specifically, in three size fractions defined by smooth collection efficiency curves with 50% cut-points at 10 µm and 3.5 µm. A personal sampler is proposed based on the inertial size separation of airborne particles and on the subsequent deposition on a 47- or 50-mm diameter membrane filter. The sampler operates with a standard rechargeable battery personal pump in a sealed circuit with a suction flow rate of 4 Lpm through the filter. Of this, 2 Lpm are recirculated into the inlet nozzle on the axis as a filtered air core of 4-mm diameter. A flow rate of 2 Lpm of external air is drawn through the surrounding ring of equal area. While proceeding through the nozzle, larger particles penetrate deeper into the clean ...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.