Abstract

The integrity of data gathered by a body worn personal radiofrequency exposure meter has been investigated using computational methods. Simulations show that the distortion of the field by the body will cause measurement errors with a traditional 3-axis dipole field sensor. A modified measurement method for body worn personal exposure meters based on an array of 2-axis dipole field sensors is proposed. The approach minimizes the impact of body shielding (shadowing of the field) and could provide information on the average internal specific absorption rate (SAR) in various parts of the body.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.