Abstract

Periprosthetic bone loss may lead to major complications in total hip arthroplasty (THA), including loosening, migration, and even fracture. This study analysed the influence of femoral implant designs on periprosthetic bone mineral density (BMD) after THA. The results of all previous published studies reporting periprosthetic femoral BMD following THA were compiled. Using these results, we compared percent changes in bone loss as a function of: femoral stem fixation, material, and geometry. The greatest bone loss was in the calcar region (Gruen Zone 7). Overall, cemented stems had more bone loss distally than noncemented stems, while noncemented stems had more proximal bone loss than cemented stems. Within noncemented stems, cobalt-chromium (CoCr) stems had nearly double the proximal bone loss compared to titanium (Ti) alloy stems. Finally, within noncemented titanium alloy group, straight stems had less bone loss than anatomical, tapered, and press-fit designs. The findings from the present study quantified percent changes in periprosthetic BMD as a function of fixation method, alloy, and stem design. While no one stem type was identified as ideal, we now have a clearer understanding of the influence of stem design on load transfer to the surrounding bone.