Abstract

The authors assess the reproducibility of retinal nerve fiber layer (RNFL) thickness measurements using scanning laser polarimetry. The right eyes of five healthy individuals were scanned by two trained operators (S.T.H., H.I.) using the Nerve Fiber Analyzer II (Laser Diagnostic Technologies, Inc., San Diego, CA, U.S.A.). Each participant was scanned six consecutive times per session for five sessions carried out over separate days within a two-week period. Intra- and interoperator measurement reproducibility was assessed. Intraoperator reproducibility was high. Mean coefficients of variation for total RNFL thickness measurements were 4.48 +/- 1.76% and 4.92 +/- 2.32% for operators 1 and 2, respectively. Interoperator reproducibility was high (p = 0.20-0.93) if a single ellipse was applied to the images of both examiners. However, measurement reproducibility worsened if each operator created his own measurement ellipse (p < or = 0.05 for 3 out of 5 participants). Interoperator error in image analysis can be minimized by using a single ellipse for the baseline and all subsequent scans even if subsequent scans are acquired by different operators.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.