Abstract

In 1991, the Dental Implant Clinical Research Group initiated a long-term clinical study in cooperation with the Department of Veterans Affairs to investigate the influence of implant design, application, and site of placement on clinical performance and crestal bone height. As part of this investigation, Periotest values for 2,212 root from implants were determined at second-stage surgery and during a 24-month follow-up period. Mean Periotest values decreased for implants placed in quality 1 and 2 bone, did not change for implants in quality 3 bone, and increased for implants in quality 4 bone. Implants in the posterior maxilla and single implants in the anterior maxilla had increasing mean Periotest values as compared with decreasing values for implants in other regions. Mean Periotest values for uncoated implants decreased gradually to approach those of hydroxyapatite-coated implants.