BackgroundWe examined whether perioperative intensive insulin therapy (IIT) using an artificial pancreas (AP) with a closed-loop glycemic control system can be used to prevent hypoglycemia in surgical patients. MethodsBetween 2006 and 2012, perioperative glycemic control using an AP was performed in 427 patients undergoing general surgery. A total of 305 patients undergoing IIT using an AP in the target blood glucose range of 80 to 110 mg/dL were enrolled in the study. Data were collected prospectively and were reviewed or analyzed retrospectively. ResultsNo patients had hypoglycemia. Perioperative mean blood glucose level and achievement rates in target blood glucose range of 80 to 110 mg/dL were 100.5 ± 11.9 mg/dL and 88.1% ± 16.0%, respectively. For the 3 primary operative methods, including hepatic, pancreatic, and esophageal resections, there were no significant differences in glycemic control stability between the types of surgery. ConclusionPerioperative IIT using an AP with a closed-loop glycemic control system can be used to prevent hypoglycemia and maintain stable glycemic control with less variability of blood glucose concentration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.