Abstract
An ultraprecision system to evaluate the periodic error of linear encoders (PEL) has been developed. Our system is based on a homodyne laser interferometer as a reference scale. Conventional studies which used an interferometer with unknown periodic error resulted in insufficient reliability because the periodic error of an interferometer (PEI) directly overlaps on the evaluated PEL. In this study, we explicitly reveal the homodyne interferometer in the proposed system has a sufficiently small PEI of less than 10 pm. This PEI is determined by analyzing the amplitude variation of interference signals, which we refer to as “self-evaluation method.” We have demonstrated the applicability of the proposed system through an actual evaluation of one of the most precise linear encoders with a PEL of sub-nanometer order.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.