Abstract
This case report presents a histologic and histomorphometric evaluation of the peri-implant tissues of a HA-coated implant retrieved due to peri-implantitis after a 10-year loading period. The implant was retrieved with a trephine and treated to obtain thin ground sections. At low-power magnification mostly compact, mature bone with small marrow spaces could be observed at the interface with the implant. The coating was always present in the areas where bone was detected, the bone was always in close contact with the coating, and there was no detachment between the metal and coating or between the coating and bone. Areas of bone remodeling were demonstrated by the presence of many secondary osteons and reversal lines close to the implant surface. The bone-implant contact percentage was 36.3% ± 1.2%. The percentage of the implant surface covered by the HA coating without bone, where bone may have detached during retrieval, was 32.6% ± 2.8%. This HA-coated implant, continued to demonstrate more than adequate BIC after many years of function and the potential to maintain osseointegration in the long term.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The International Journal of Oral & Maxillofacial Implants
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.