Abstract

ObjectiveIn patients with non-valvular atrial fibrillation (NVAF), the left atrial appendage occluder (LAAO) is an alternative treatment for stroke prevention. However, thromboembolic event still occur, and the predictors are unknown. MethodsThe first Asian long-term follow-up study consisted of 308 patients with mean age 71.9±9.5 years, mean CHA2DS2-VASc 4.1 ± 1.8 since 2013. Primary outcome was defined as any type of ischemic stroke/transient ischemic attack (TIA), systemic embolization and cardiovascular death. ResultsThere was no procedural-related TIA or stroke. After a mean follow-up of 38±16 months, the ischemic stroke/TIA rate was 1.9 and cardiovascular death rate 0.3 per 100 patient-year. The rate of peri-device leak (PDL) was 11.9% and device-related thrombus (DRT) 2.6%. In the multivariable analyses, PDL was the only independent predictor of stroke/TIA (hazard ratio 5.5, p=0.008). CHA2DS2-VASc score, prior history of stroke, DRT and post-procedural anti-thrombotic regimen/duration were not associated with outcomes. Implantation of Watchman was associated with PDL (odds ratio 4.35, p=0.001). ConclusionsPDL is the only independent predictor of post-LAAO stroke. The risk of stroke for patients with NVAF may be controllable after LAA is occluded, because PDL is preventable and treatable.