Abstract
The silica aerogel threshold Cherenkov counter for the Jefferson Lab Hall A spectrometer was tested at CERN with a mixed beam of positrons, pions, and protons in the 1–4 GeV/ c momentum range. Aerogel from Airglass, Sweden, with a refractive index of 1.025 was used. For this momentum range and refractive index only positrons and pions are above the Cherenkov threshold. Detection efficiencies of 96% for pions and of 98% for positrons were measured. A 5% fake proton detection was also measured, mostly caused by induced δ-rays and accidentals generated by high counting rates.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.