Abstract
Objective To verify the main analyzing performance of the 7600 automatic biochemical analyzer,production of Hitachi Corporation of Japan.Methods Following the method how EP5A,EP6A were tested as described in the instruction of the National Committee for Clinical Laboratory Standards (NCCLS),precision,linearity and reportable range of the selected biochemical testing items were evaluated.Results All the withinrun precision of Hitachi 7600 testing items were within the required 1/4 CLIA ′88 but Na,all the betweenrun precision were within the required 1/3 CLIA ′88 but Na;the linear verification results got the regression equation of the theoretical and measured values y=bx+a,in which b was within the range of 0.971.03,a was within an acceptable range;the reportable range verification results showed that after the samples being diluted by different proportions,the theoretical/measured values were all between 90% and 110%,indicating that within a certain range of sample dilution the test results were reliable.Conclusions Precision,linearity and reportable range of the detection system met the basic requirements of the experiments in clinical diagnosis and the method detecting Na should be improved.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.