Abstract

In this paper, performance tests of a hearing aid utilized intensely by a person living a hearing trouble as stated in the methods described in IEC 60118 Electroacoustic -Hearing Aids series standards; such as total harmonic distortion, automatic gain control, output sound pressure level (OSPL), high frequency averaging, saturation sound pressure level (SSPL) and battery current measurement, have been made by applying hearing aids performance tests system infrastructure equipped in TUBITAK UME Medical Metrology Laboratory. The performance test parameters of the measured hearing aid have been clarified distinctly according to measurement outcomes and attention is put on the importance of effective market surveillance and/or control of hearing instruments in particular.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.