Performance test of a Ti/Au bilayer TES microcalorimeter has been made in combination with a continuous adiabatic demagnetization refrigerator (CADR). The CADR has four stages of ADR to produce continuous cooling by recycling them in dedicated order, and is cryogen‐free utilizing a 4K‐GM refrigerator. We installed a Ti/Au bilayer TES microcalorimeter and 420‐series SQUID array to readout the X‐ray signal on the 1st (coldest) stage of the CADR. We successfully operated the CADR at temperature of 120 mK in continuous mode more than 27 hr, however, FWHM energy resolution of the TES microcalorimeter was degraded to 45 eV at 6 keV, as compared to 10 eV when measured in a dilution refrigerator. This is mainly because the temperature stability was not good enough (about 0.6 mK) and the operation temperature was not sufficiently lower than the transition temperature Tc = 135mK of the TES. We operated the TES microcalorimeter at the operation temperature of 105 mK in one‐shot mode and the resolution was improved to 30 eV. We also found that the operating point of the TES was affected by the magnetic field of the 3rd and 4th ADR recycle. More complete shielding of the magnetic field is essential for further improvement of the performance of the TES microcalorimeter.