Abstract

We have developed a miniature gamma-ray endoscopic probe consisting of dual BGO detector probes for tumor detection inside the body cavities. The dual detector system was coupled with random coincidence to decrease the distant background radiation and to improve its spatial resolution for tumor localization. The performance of the probe was investigated with a point source and a water phantom. A solution of positron emitting 18F isotope was used as the source. Clinical trials of the probe were done to localize tumors on the skin surface of four subjects carrying tumors close to the body surfaces, into whom 67Ga-citrate and 18F-FDG radiopharmaceuticals were injected. Measurements indicated that the spatial resolution of the dual detector probes is around 1.5 times better than the single detector probe, and both single and dual detector endoscopic probe systems are capable of localizing a tumor on a large photon background. The endoscopic probe may be easier to insert inside body cavities due to the small crystal size and the flexible light guides. A single detector probe with higher sensitivity may be useful in searching for tumors over a wide intracavity area but a dual detector probe can be used for precise tumor localization. The detector probe may also be suitable for intraoperative observation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.