Abstract

Maximum limits of 0.02% are recommended for precision of IR analysis for fat, protein, and lactose, and 0.04% for total solids. Recommended maximum systematic errors are 0.06% for fat, protein, and lactose, and 0.12% for total solids, when instrument results are compared with results of specified AOAC methods. Recommended maximum mean differences between instrument and standard results are based on 95% confidence limits for the mean of 8 samples; precision and systematic errors are calculated as the standard deviation of difference between duplicates for 8 samples, and systematic errors are calculated as the standard deviation of difference between instrument and reference results for 8 samples. Automated sections for milk of Official Methods of Analysis were rewritten to eliminate references to individual instruments, to eliminate possible differences between calibration procedures for different instruments, and to stress important instrument operational characteristics which can have significant effects on instrument performance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.