Abstract

The less expensive and high resolution Multi-gap Resistive Plate Chamber (MRPC) opens up a new possibility to find an efficient alternative detector for the Time of Flight (TOF) based Positron Emission Tomography, where the sensitivity of the system depends largely on the time resolution of the detector. In a layered structure, suitable converters can be used to increase the photon detection efficiency. In this work, we perform a detailed GEANT4 simulation to optimize the converter thickness towards improving the efficiency of photon conversion. A Monte Carlo based procedure has been developed to simulate the time resolution of the MRPC-based system, making it possible to simulate its response for PET imaging application. The results of the test of a six-gap MRPC, operating in avalanche mode, with 22Na source have been discussed.