Abstract
We have previously demonstrated a novel interferometric multispeckle Fourier domain diffuse correlation spectroscopy system that makes use of holographic camera-based detection, and which is capable of making in vivo pulsatile flow measurements. In this work, we report on a systematic characterisation of the signal-to-noise ratio performance of our system. This includes demonstration and elimination of laser mode hopping, and correction for the instrument’s modulation transfer function to ensure faithful reconstruction of measured intensity profiles. We also demonstrate a singular value decomposition approach to ensure that spatiotemporally correlated experimental noise sources do not limit optimal signal-to-noise ratio performance. Finally, we present a novel multispeckle denoising algorithm that allows our instrument to achieve a signal-to-noise ratio gain that is equal to the square root of the number of detected speckles, whilst detecting up to ∼1290 speckles in parallel. The signal-to-noise ratio gain of 36 that we report is a significant step toward mitigating the trade-off that exists between signal-to-noise ratio and imaging depth in diffuse correlation spectroscopy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.