To meet the testing demands and overcome supply chain issues during the SARS-CoV-2 pandemic, many clinical laboratories validated multiple SARS-CoV-2 molecular testing platforms. Here, we compare three different molecular assays for SARS-CoV-2 that received emergency use authorization (EUA) from the U.S. Food and Drug Administration. In order to determine the agreement among Roche cobas® SARS-CoV-2 Test (Cobas), Abbott RealTime SARS-CoV-2 assay (ART), and Mayo Clinic Laboratory SARS-CoV-2 Molecular Detection Assay (Mayo LDT), 100 each of anterior nares (AN), nasopharyngeal (NP), oropharyngeal (OP), and NP+OP swabs were tested on each platform. The consensus result was defined as agreement by 2 or more methods. Furthermore, 30 positive NP swabs from each molecular platform (n = 90 total) were tested on the three platforms to determine the PPA among positive samples. ART platform called more specimens positive than the other two platforms. All three assays performed with greater than 90% agreement for NP specimens throughout the study.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.