Abstract

The performance of liquid chromatography-isotope ratio mass spectrometry (LC-IRMS) for polar halogenated compounds was evaluated. Oxidation capacity of the system was tested with halogenated acetic acids and halogenated aromatic compounds. Acetic acid (AA) was selected as a reference compound for complete oxidation and compared on the molar basis to the oxidation of other analytes. The isotope values were proofed with calibrated δ(13)C values obtained with an elemental analyzer (EA). Correct isotope values were obtained for mono- and dichlorinated, fluorinated, and tribrominated acetic acids and also for aniline, phenol, benzene, bromobenzene, chlorobenzene, 1,2-dichlorobenzene, 2,4,6-trichlorophenol, pentafluorophenol, and nitrobenzene. Incomplete oxidation of trichloroacetic acid (TCA) and trifluoroacetic acid (TFA) resulted in lower recovery compared to AA (37% and 24%, respectively) and in isotopic shift compared to values obtained with EA (TCA Δδ(13)C(EA/LC-IRMS) = 8.8‰, TFA Δδ(13)C(EA/LC-IRMS) = 6.0‰). Improvement of oxidation by longer reaction time in the reactor and increase in the concentration of sulfate radicals did not lead to complete combustion of TCA and TFA needed for δ(13)C analysis. To the best of our knowledge, this is the first time such highly chlorinated compounds were studied with the LC-IRMS system. This work provides information for method development of LC-IRMS methods for halogenated contaminants that are known as potential threats to public health and the environment.