Abstract

We evaluated the performance of the Vitek MS for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories. With a total of 424 well-characterized isolates, the results of the Vitek MS were compared to those of conventional methods and 16S rRNA gene sequencing. The Vitek MS correctly identified 97.9 % of the isolates tested to species level. The Vitek MS correctly identified the species of 97.2 % of the staphylococci (95.9 % of coagulase-negative staphylococci), 97.8 % of the streptococci, and 100 % of the enterococci. For the identification of Gram-positive cocci isolates, the overall concordance rate between conventional identification and the Vitek MS was 94.5 %. The Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system can be a reliable and rapid method for the identification of most relevant Gram-positive cocci. In addition, expanding the database of the Vitek MS, especially for coagulase-negative staphylococci, is needed to enhance the performance of the Vitek MS.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.