To evaluate the use of the Spot Vision Screener (Spot; Welch Allyn, Skaneateles Falls, New York, USA) for detection of amblyopia risk factors in children aged 6months to 3 years, as defined by the 2013 guidelines of the American Association for Pediatric Ophthalmology and Strabismus. Reliability analysis. In this study, children seen from June 1, 2012, to April 30, 2016 were tested with the Spot during a routine visit. Enrolled children underwent a comprehensive eye examination including cycloplegic refraction and sensorimotor testing within 6months of the testing date by a pediatric ophthalmologist masked to the Spot results. A total of 184 children were included. The Spot successfully obtained readings in 89.7% of patients. Compared with the ophthalmologist's examination, the Spot had an overall sensitivity of 89.8% and a specificity of 70.4%. The Spot achieved good sensitivity and specificity for detection of amblyopia risk factors in this young cohort, particularly in the older subgroup. Our data offer support for automated vision screening in young children.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.