Abstract

Amblyopia is a common cause of monocular vision impairment and disproportionally affects developmentally delayed children. Photoscreeners have been suggested as a method to detect amblyopia risk factors (ARFs) in children with developmental disabilities who may not be amenable to traditional vision screening methods. The Spot Vision Screener is a commonly used photoscreener for detecting ARF and has shown excellent sensitivity and accuracy in the general pediatric population. The purpose of this study was to evaluate its accuracy in children with Down syndrome and other special needs. Children with various disabilities or delays were recruited from outpatient clinics at the Children's Hospital of Colorado. Participants had their photograph taken with Spot before and after pupil dilation and cycloplegia. Images were compared to results of a comprehensive clinical eye examination. A total of 100 children participated in the study. Images could not be obtained in 12 children; 5 children did not attend their clinical examination. The overall sensitivity of Spot was 90%, with a positive predictive value of 80% in undilated subjects. The area under the receiver operator curve (AUROC) was 0.68 (95% CI, 0.57-0.79), which did not differ significantly from the AUROC after dilation/cycloplegia (0.68; 95% CI, 0.54-0.81). The Spot Vision Screener could be used by primary care clinics and vision screening programs with a high sensitivity to determine which patients with Down syndrome and special needs should be referred for clinical examination.